Subject: S2 Data Quality: byte swap error
S2 Data Quality: byte swap error:
Here is an email from Rolf describing the recently found byte swap error
that affects H1 IOO PSL, SUS coil readbacks and PEM channels acossiated
with H1.
> http://blue.ligo-wa.caltech.edu/ilog/pub/ilog.cgi?group=detector&date_to_view=07/25/2003&anchor_to_scroll_to=2003:07:25:17:18:39-rick
This was happening during S2.
Daniel
>During signal injection testing last week, a problem was noted with
data from
>two ADCU (h1adcusus and h1adcupem). This was subsequently tracked
down to a
>software problem in these two ADCU. These two ADCU are different than
all others
>in LIGO in that they are the only two directly connected to the framebuilder
network.
>
>The problem originated when the framebuilder network at Hanford was
switched
>over to a newer high speed network back in January. The new net was
installed to
>accomodate the high data load. The older network cards had hardware
byte
>swapping (needed to send data from a Pentium to the Sun), but the
new network
>boards do not have this feature. Therefore, the software was changed
in these
>ADCU to do the byte swapping. Unfortunately, it was not noted that
while the
>ADCU writes the data as shorts, the framebuilder, using 32 bit DMA
to recieve
>the data, actually reads integers. This results in inproper byte swapping.
>
>The net result in the stored data frames is that data samples for
these two ADCU
>are correct in value but out of time order. So the time series, from
the
>beginning of a frame, instead of being (S)ample 0, S1, S2, S3, S4,
..., got
>stored as S1,S0,S3,S2,S5,S4 .... ie pairs of data samples are swapped
in the
>time series.
>
>The software was corrected, reloaded and tested yesterday. Unfortunately,
the S2
>data must somehow be corrected or at least be noted that this problem
exists in
>that data.
>
>-rolf